The mechanism behind my GeSe glucose sensor is optical rotation. Glucose is chiral, it turns the plane of linearly polarized light, and the detector converts that turn into a change in photocurrent. What I want to establish next is how much of the signal I measure is that chirality-specific effect, and how much comes from other things a glucose solution does to a beam. This post sets out the arithmetic that frames the question and the control experiment I plan to use to answer it.
How big is the physical effect?
The optical rotation of a solution is α = [α] · l · c, where [α] is the specific rotation, l is the path length in decimetres and c is the concentration in grams per millilitre. For D-glucose at equilibrium, [α] is about +52.7° per dm per (g/mL) at the sodium D line, and it falls as the wavelength gets longer, roughly as 1/λ². For an illustrative estimate at red wavelengths I take a round 40°, which is good to 20% or so and is all this arithmetic needs.
Now assume a cell 1 cm long, which is 0.1 dm, and a concentration of 100 mg/dL, which is 1 g/L or 0.001 g/mL.
α ≈ 40 × 0.1 × 0.001 ≈ 4 × 10⁻³ degrees.
Under those assumptions a 1 cm path of 100 mg/dL glucose rotates the plane of polarization by a few thousandths of a degree, about 4 × 10⁻⁵ degrees for each mg/dL. Across a calibration range of 50 to 300 mg/dL, the rotation in a 1 cm cell changes by about 0.01°. That is small enough that the closed-loop polarimeters of the Coté group (Cameron and Coté, 1997, doi:10.1109/10.649993) rely on modulation and phase-sensitive detection to resolve it. Both the path length and the specific rotation are assumptions in this estimate, so the result is an order of magnitude and not a prediction for my cell.

In my setup the solution sits in a small vial between the half-wave plate and the detector, so the path is centimetres and not metres.
What the detector reports
The quantity I read out is not the rotation itself. I step a half-wave plate through a half-turn, record the photocurrent at each step, fit a sinusoid, and report the phase of that sinusoid. The plot below, from my 2024 paper, shows that fitted phase against glucose concentration. It rises steadily across the prepared solutions.

A fitted phase is a calibration coordinate. It tracks concentration, which is what a calibration needs, and it sits downstream of several steps: the rotation of the beam, any change in how the beam is steered or focused, the response of the flake, and the fit. The research question is the transfer function between the physical rotation and the phase I report. A lock-in polarimeter reports rotation in degrees directly. Mine reports a phase, and relating the two quantitatively needs an independent measurement.
What else could move the phase
A glucose solution changes more than the polarization of light, and listing the candidates is the first step in designing a control.
The refractive index of the solution rises with glucose concentration, by something of order 10⁻⁴ for each 100 mg/dL. A cylindrical vial acts as a weak lens, so a change in index changes how the beam is focused or steered, and my flake is small, so a slightly displaced spot can change how much light it collects. The vial walls can carry stress birefringence that varies from one fill to the next. A diode laser’s polarization state can drift with temperature, which matters because one phase value requires a full scan of the half-wave plate, and the measurement takes time.
The last of those is the cheapest to test. If solutions are measured in ascending order of concentration, any steady drift in the laser or the alignment during a session produces a trend in the fitted phase that looks like a calibration slope. The fix is to randomize the order of the solutions and to put a water reference between every few of them. If the slope survives, drift is out of the picture.
One more effect belongs to the rotation itself and not to the confounds. Glucose in solution mutarotates. Freshly dissolved crystalline α-D-glucose has a specific rotation near +112°, and over a few hours, as the α and β forms interconvert, it settles at the equilibrium value of +52.7°. A solution measured minutes after it was made therefore has a rotation that is still changing, and the speed of mutarotation depends on temperature and pH. That matters for a design that compares solutions made at different times. I would prepare all solutions together, let them equilibrate for a fixed period, and record the time of every scan so that any residual trend can be checked against the clock.
The control I plan to run
The drawing below summarizes the design. All three solutions go through the same cell, with the same path length, at the same concentrations in mg/dL.

L-glucose is the mirror image of D-glucose. It has the same refractive index, the same density and the same osmotic behaviour in solution, and it rotates the plane of polarization by the same amount in the opposite direction. It is commercially available. If the fitted phase moves up with D-glucose and down with L-glucose, the signal is sensitive to the handedness of the molecule, which only optical rotation can explain. If both sugars push the phase the same way, that part of the signal comes from something they share: index, density or osmotic effects on the cell and the beam. A third solution, an index-matched solute with no optical rotation, gives a second view of the same question.
Three further checks would sharpen the answer. The first is path length. The rotation is proportional to the product of concentration and path length, so doubling the cell length should double the effect for a given concentration. A lensing or steering effect has no reason to scale that way. The second is wavelength. Optical rotation grows toward shorter wavelengths roughly as 1/λ², so a second laser would give a rotation signal a predictable change in size. The third is a reference measurement. A conventional polarimeter on the same cell with the same solutions would give the physical rotation to compare against the phase my detector reports. At a few thousandths of a degree per 100 mg/dL in a 1 cm cell it is a demanding measurement and would likely need modulation, but even an upper bound would be informative.
What each outcome would teach
A sign reversal between D- and L-glucose, together with scaling with path length, would show that the chirality-specific mechanism accounts for the signal. The next question would then be how a small GeSe flake turns a rotation of a few thousandths of a degree into a measurable phase shift, and that is a question about the detector’s optics and its polarization-dependent absorption, which the anisotropy discussion in the first post of this series points toward.
If the phase moves the same way for both enantiomers, the signal contains a common component, most likely tied to refractive index, and the work shifts to separating the two: a reference channel, a second wavelength, or a model trained to use the features that differ between them. In a mixture such as a biological sample, many solutes change refractive index, so quantifying the chirality-specific fraction is the step that decides how specific a polarimetric signal can be.
A mixed result, with some reversal and some shift common to both sugars, seems the most likely to me and the most useful. It would give the fraction of the signal that is chirality-specific and show how much of the machine-learning model’s job is to handle the rest.
The data I have show that the fitted phase follows concentration in solutions measured on my setup. The control experiment determines how much of that is rotation. VitaSensor remains laboratory research, at the stage of a third-generation prototype, and nothing in this post concerns measurements outside the laboratory.
Sources and links
- Paper: Kopp, M. B. (2026). Next-generation polarimetric biosensors: machine learning–driven GeSe photodetectors for noninvasive glucose monitoring. Biosensors and Bioelectronics: X, 28, 100725. doi.org/10.1016/j.biosx.2025.100725 (open access, CC BY 4.0).
- Earlier paper: Kopp, M. B. (2024). Flexible nanomaterial sensors for non-invasive health monitoring. Applied and Computational Engineering, 72, 45–58. doi.org/10.54254/2755-2721/72/20240981.
- Cameron, B. D., and Coté, G. L. (1997). Noninvasive glucose sensing utilizing a digital closed-loop polarimetric approach. IEEE Transactions on Biomedical Engineering, 44(12). doi.org/10.1109/10.649993.
- Google Scholar and GitHub; project repository gese-glucose-biosensor.
- Related posts: how GeSe reads polarization without a polarizer, gradient boosting on the photocurrent signal, polarizer-free polarimetry versus bench instruments.